Main Technique Specifications
| Function Model | CL-6850 |
| Range | Free chlorine :0-20.00 mg/L(ppm), Hypochlorous acid (HOCl) :0-10.00 mg/L(ppm), PH value :0-14pH, Temperature: 0-60 ℃ |
| Accuracy | Free chlorine : ±1% or ±0.01 mg/L, Hypochlorous acid (HOCl) :±1% o ±0.01 mg/L, PH value :±0.02pH , Temperature: ±0.5℃ |
| Temp. Comp. | Manual/automatic pH compensation feature(0-14) and temperature compensation(0~60℃) |
| Electrode | Free chlorine Sensor,PH sensor |
| Display | LCD display |
| Operation Temp. | 0~60℃ |
| Current Output | Isolation 4~20mA |
| Control Output | ON/OFF High limit,low limit relay |
| Power | AC 110/220V±10% 50/60Hz, DC 24V, DC 12v |
| Working Environment | Ambient Temp. 0~50℃, Relative Humidity ≤90% |
| Overall Dimensions | 96×96× 115mm(H×W × D) |
| Hole Size | 91 ×91mm(H×W) |
| Installation Mode | Panel mounted(embedded) |
| Protection Grade | IP 65 |
product details
Application